The HD32MT.1 is a 16-channel data logger capable of capturing and logging the values measured by a series of sensors connected to its inputs.
The data logger is completely programmable by the user and is therefore very versatile. The supplied HD32MTLogger application software, supplied with the instrument, allows simple and intuitive programming by using graphic interfaces, without the need of learning any programming
language, thus minimizing the time needed to make the system operational.
The values recorded by the instrument can be transferred to a PC by using the HD32MTLogger software. The data logger can be configured to memorize the instant value, the minimum value, the maximum value, the average value and the standard deviation of the measurements.
For measurements that require the counting of pulses, the total counted pulses can be stored.
Technical characteristics
Sizes 222x140x63 mm
Weight About 1 kg
Case material Coated aluminium
Operating conditions -20 … 50° C, 0 … 85% RH no condensation
Storage temperature -25 … 65° C
Power Supply 12 … 30 Vdc
Absorption 40mA @ 12 Vdc
Data acquisition interval from
sensors Programmable from 1 to 60 seconds
Data logging interval Programmable from 2 seconds to 24 hours
Storage capacity 4 MB internal memory
SD memory card reader up to 2 GB
Number of samples that can be stored
The storage of a record consisting of N values requires ( 4x N) bytes of memory plus 8 bytes for the date and time.
Analog inputs
8 channels, each channel can be used as a differential input or alternatively as 2 single-ended inputs.
Measurement ranges:
± 25mV, ± 100mV, ± 1000mV, ± 2500mV
Resolution: 16 bit, Accuracy: 0.01% f.s.
Input impedance: 100Mohm
Digital input/ output ports ( I/ O)
8 ports, each configurable as an input for connecting a sensor or alarm output.
TTL logic levels ( 0Ò Vin3V)
Max. input voltage: 5.5 V
Inputs for high frequency pulse counting
2 inputs
Frequency of pulses 50kHz max.
TTL logic levels ( 0Ò Vin3V)
Minimum pulse duration 10 ¼ s
Inputs for number of potential free
contact opening/ closing counting
2 insulated inputs
Switch frequency 50Hz max.
Minimum opening or closing time 10 ms
RS485 connection 1 RS485 port for connection of anemometers HD2003 and
HD52.3D… series
RS232 connection 2 RS232 ports, one for connection to PC or to optional
Radio Modem and one for connection to optional GSM module.
Sub-D 9-pole male connectors
Alarm outputs
2 insulated voltage-free contact outputs
Contact: max. 1A @ 30Vdc resistive load
You can configure the single digital I/ O ports as alarm outputs
Auxiliary supply outputs
+ 5V regulated, max. 500 mA
+ Vsw ( switched) : with same value of the power input, it is active only during acquisition of measurements